Temperature Mapping Requirements: UKAS Accredited vs Traceable Mapping
What the MHRA and GDP require for temperature mapping, why it should be risk based, and how UKAS accredited mapping differs from traceable.
The MHRA requires storage areas to be temperature mapped, but leaves the method to you. That makes the quality of the mapping, and the evidence behind it, your responsibility. With traceable mapping, nobody independent audits the provider, so you cannot know what you are getting.
The MHRA says you must map. It does not say how.
If you store medicines, vaccines or other temperature sensitive products, the MHRA expects your storage areas to be temperature mapped. Auditors will ask to see the evidence. What the MHRA does not do is tell you what that mapping should look like: how many sensors, where they go, how long they run, how accurate they must be, or what the final report should contain.
That gap is why two mapping studies of the same warehouse can look completely different, and why quotes for “temperature mapping” can vary so widely. Some studies will stand up to an inspector’s questions. Others are little more than a set of logger downloads in a spreadsheet.
In this guide we explain what the regulations actually ask for, why mapping should be a risk based exercise, and the difference between UKAS accredited and non accredited mapping, including the part most people never ask about: measurement uncertainty.
What the regulations actually say
The core requirement sits in Chapter 3.2.1 of the Good Distribution Practice (GDP) Guidelines, which the MHRA enforces in the UK. It asks for an initial temperature mapping exercise on the storage area before use, under representative conditions. Permanent monitoring equipment should then be positioned according to the results, in the areas that see the extremes of temperature. The exercise should be repeated based on a risk assessment, or whenever significant changes are made to the facility or its temperature control equipment.
The MHRA Inspectorate has added some practical expectations in its own guidance:
- Map before stock is stored where possible, and repeat once the area is fully stocked.
- Document the data and a risk assessment that identifies any hot or cold spots.
- Use that risk assessment to decide where the permanent monitoring probes go.
- Repeat the exercise to cover seasonal variation (in practice, summer and winter) and after any reconfiguration.
- Justify the length of the mapping exercise through a risk assessment of the storage area.
Notice what is missing. There is no sensor count, no grid spacing, no accuracy requirement for the sensors and no requirement for the mapping to be accredited. Every one of those decisions is left to you, and you must be able to justify each of them to an inspector.
Because the regulator does not define a method, we base our process on the World Health Organization’s Temperature mapping of storage areas (Technical Report Series No. 961, Annex 9). It is one of the most widely recognised references for how mapping should be planned, carried out and reported.
Mapping is a risk based exercise
There is no single “correct” mapping study. The right level of mapping depends on what you store, how you store it and what could go wrong. A 2–8 °C vaccine fridge, an ambient pharmaceutical warehouse and a chilled food store each carry very different risks.
That is why we start every project with a conversation, not a price list. We work with you to identify the risks that matter in your area, such as:
- the products stored and their temperature limits
- the size of the area and the height goods are stacked to
- doors, loading bays, windows, roof spaces and external walls
- the position of heating, cooling and air handling units
- how the area is used day to day, including weekends and shutdowns
- your audit body’s expectations and your quality system
Those risks then drive the decisions: how many sensors, where they go, how long the study runs, whether it needs repeating seasonally, and which documents you need at the end. A small clinic fridge may need a simple study over 24 to 48 hours. A multi zone GDP warehouse may need well over 100 sensors running for a week or more, in both summer and winter.
The aim is mapping that is proportionate to your risk: enough to stand up to an inspection, without paying for work you do not need.
Accredited vs non accredited mapping
Most mapping on the market is described as “UKAS traceable”. Very little is UKAS accredited. The two terms sound similar but mean very different things.
UKAS traceable mapping means the equipment can be linked back, somewhere along a chain, to a UKAS calibrated reference. It says nothing about how the mapping itself was planned, carried out, analysed or reported. Nobody independent has audited the mapping process, the staff doing it, or the way the results are calculated.
UKAS accredited mapping means the mapping activity itself sits within the provider’s ISO/IEC 17025 scope of accreditation. UKAS independently assesses the whole process: the method, the competence of the engineers, the calibration of every sensor, the calculation of uncertainty and the content of the report. The provider is reassessed regularly and must keep meeting the standard to keep its accreditation.
TekTroniks is UKAS accredited to ISO/IEC 17025:2017 for temperature mapping across a range of −45 °C to +30 °C. You can check this on our UKAS Schedule of Accreditation.
Traceable mapping is often cheaper. But when an inspector asks how you know your mapping results are correct, “the loggers were traceable” is a much weaker answer than “the study was carried out under an independently audited, accredited process, and here is the certificate”.
With traceable mapping, you do not know what you are getting
This is the point that matters most. Nobody independent audits a company that offers traceable mapping. How it plans a study, calibrates its loggers, analyses the data and writes the report is checked by no one but itself.
Even the word traceable leaves the key questions unanswered. Traceable to what? Was the reference probe calibrated by an accredited laboratory, and is it still in date? How were the loggers compared with it: in a controlled bath or chamber, at which temperatures, and with what uncertainty? A traceable mapping report rarely says. Without those answers, the accuracy of every reading in the study is assumed, not known. Every supplier applies the word in its own way, so two traceable certificates can mean completely different things. Examine any traceable claim closely and it soon falls apart, unless the work was done to a recognised standard. ISO/IEC 17025 is that standard, and UKAS accreditation is the independent proof that it has been met.
That provider may be excellent, or it may be poor. From the report alone, you cannot tell which. The only way to find out is to audit the provider yourself: its methods, its sensor calibration, its data handling and the competence of its engineers, with the technical knowledge to judge them. Few customers have the time or expertise to do that.
If you do not, the risk stays with you. When an inspector questions the mapping, it is your quality system that has to answer for it. UKAS accreditation means that audit has already been done, independently, against ISO/IEC 17025.
How the mapping sensors are calibrated
A mapping study is only as good as the sensors used to carry it out. This is where the difference between providers is often greatest, and least visible.
With non accredited mapping, the loggers are commonly checked in house against a single “traceable” reference probe. The reference probe may hold a calibration certificate, but the in house check usually does not. Typically there is no documented uncertainty for the check, no control of the bath or chamber used, and no independent oversight of how it was done. The loggers are then described as traceable, even though the link between them and the reference is undocumented.
With accredited mapping, every sensor used in the study is calibrated under our ISO/IEC 17025 accreditation and carries its own UKAS calibration certificate. That calibration follows the same controls described in our guide to UKAS accredited vs traceable calibration: mapped baths and chambers, two reference thermometers on every calibration, regular checks of reference equipment and a full uncertainty budget.
The result is that the error and uncertainty of each mapping sensor are known, documented and independently verified, rather than assumed.
Mapping uncertainty: the question most people never ask
Every measurement has an uncertainty: a range either side of the reading within which the true value is expected to lie. For a mapping study, that uncertainty tells you how much confidence you can place in every hot spot, cold spot and pass or fail result in the report.
As an accredited laboratory, we calculate an uncertainty budget for the mapping itself. It combines all the significant sources of error, including:
- the calibration uncertainty of each mapping sensor
- the uncertainty of the reference thermometers used to calibrate them
- sensor drift between calibrations
- the resolution of the logging equipment
- the stability and uniformity of the baths and chambers used for calibration
Each contribution is converted to a standard uncertainty and combined, then multiplied by a coverage factor to give an expanded uncertainty at a confidence level of approximately 95%:
U = k × √(u₁² + u₂² + u₃² + … + uₙ²), where k = 2
Why does this matter? Imagine a 2–8 °C cold room where one sensor records a maximum of 7.8 °C. If the mapping uncertainty is ±0.2 °C, you can be confident that location stayed within limits. If the uncertainty is ±0.5 °C, or simply unknown, you cannot. The same reading could be a pass or a hidden excursion.
With most non accredited mapping, no overall uncertainty is calculated at all. The loggers may be traceable, but nobody can tell you how accurate the mapping results actually are.
Our mapping report is a calibration certificate
This is the biggest single difference between our mapping and most of what is on the market. Because the study is carried out under our ISO/IEC 17025 accreditation, the mapping report is issued as a UKAS calibration certificate for the area mapped, bearing the UKAS mark.
A non accredited report is a document written by the provider. A UKAS certificate is a document produced under a process that has been independently assessed by the UK’s national accreditation body. For an auditor, that is the difference between taking the provider’s word for it and having independent assurance.
Our certificate and report include:
- a schematic or floor plan of the area, showing every sensor position
- minimum, maximum and average temperature for each sensor
- the data in both tabular and graphical form
- excursion analysis against your temperature limits
- Mean Kinetic Temperature (MKT), where required
- the stated measurement uncertainty for the mapping
- a UKAS calibration certificate for each sensor used
- the raw data at one minute intervals, supplied as CSV files
We also present the findings in person, and advise on where your permanent monitoring probes should be placed.
From scope to certificate: the full document trail
An inspector will rarely look at the final report alone. They want to see that the mapping was planned, justified and carried out as agreed. Our accredited process produces a complete, auditable trail of documents:
- Mapping scope. After a site visit, we agree a written scope with you. It defines the area, the storage envelope, the number of sensors, the mapping period and the acceptance limits. It is signed by both parties before any work starts.
- Map of the area. We produce a drawing of the area mapped, showing racking, doors, cooling units and the exact position and height of every sensor.
- Sensor placement justification. Every sensor position can be justified. We document why each location was chosen, based on the grid approach in the WHO guidance and the specific risks in your area, such as doors, loading bays, external walls and air handling units.
- Mapping report and UKAS certificate. The results, analysis, uncertainty and conclusions, issued as a UKAS calibration certificate.
- Monitoring recommendations. Guidance on where your permanent monitoring probes should go, based on the hot and cold spots found.
The documents you need depend on your risk and your auditor. Some customers need the full suite for a GDP inspection. Others need a lighter set for an internal quality system. We agree this with you at the start, so you get the documentation your audit requires.
Accredited vs traceable at a glance
| Typical traceable mapping | TekTroniks UKAS accredited mapping | |
|---|---|---|
| Independent audit of the mapping process | No | Yes, by UKAS to ISO/IEC 17025 |
| Sensor calibration | Often checked in house against one traceable probe | Every sensor calibrated under accreditation, with its own UKAS certificate |
| Mapping uncertainty | Rarely calculated | Full uncertainty budget, stated on the certificate |
| Final document | Provider’s report | UKAS calibration certificate plus full report |
| Agreed, signed scope | Varies | Yes, for every mapping |
| Map of the area and sensor positions | Varies | Yes |
| Justification for each sensor position | Rarely | Yes |
| Level of mapping | Often a fixed package | Risk based, agreed with you |
Questions to ask your mapping provider
- Is your mapping within your UKAS scope of accreditation, or are your loggers just traceable?
- How are your mapping sensors calibrated, and does each one have a UKAS certificate?
- What is the overall uncertainty of the mapping, and is it stated on the report?
- Can you justify the position of every sensor?
- Will I receive a signed scope, a map of the area and a UKAS certificate for the study?
Mapping you can defend
The MHRA requires temperature mapping but leaves the method to you. That freedom also puts the responsibility on you to show that your mapping was planned properly, carried out competently and measured accurately.
UKAS accredited mapping gives you that evidence: an independently audited process, calibrated sensors with known uncertainty, a full set of justified documents and a UKAS certificate for the study. Mapping that is proportionate to your risk, and that you can defend at inspection.
To discuss your mapping requirements, call us on 0115 989 0090, email [email protected] or get in touch through our website. See our temperature mapping services.
Temperature mapping FAQs
Does the MHRA require temperature mapping?
Yes. Chapter 3.2.1 of the GDP Guidelines, enforced by the MHRA in the UK, asks for an initial temperature mapping exercise on the storage area before use, under representative conditions, repeated based on a risk assessment or after significant changes.
How many sensors does a mapping study need?
The regulations do not set a number. It depends on your risk assessment: a small clinic fridge may need a simple study over 24 to 48 hours, while a multi zone GDP warehouse may need well over 100 sensors running for a week or more.
What is the difference between UKAS accredited and UKAS traceable mapping?
Traceable mapping only means the equipment links back to a UKAS calibrated reference. Accredited mapping means the whole mapping process sits within the provider’s ISO/IEC 17025 scope and is independently assessed by UKAS.
Is traceable mapping always worse?
Not necessarily. Some providers do good work. But nobody independent audits them, so you cannot know which you are getting unless you audit them yourself.
What do I receive at the end of a mapping study?
A UKAS calibration certificate and full report, including sensor positions, results, excursion analysis, the stated mapping uncertainty, a UKAS certificate for each sensor and the raw data as CSV files.
Sources
- MHRA Inspectorate: Temperature mapping, an introduction
- European Commission: GDP Guidelines questions and answers (Guidelines 2013/C 343/01, Chapter 3.2.1)